Products
Products
home
MMS6410

Core Functionality:
A 4-channel industrial monitoring system designed for comprehensive machinery protection, measuring:
✓ Vibration
✓ Displacement
✓ Speed
✓ Temperature (optional)

Key Technical Specs:
• Inputs: Accepts signals from eddy current probes, velocity sensors, tachometers, and RTD/TCs
• Measurement Ranges:

Vibration: 0-500 µm

Displacement: 0-2.0 mm

Speed: 0-20,000 RPM
• Outputs: 4x 4-20mA + 8 alarm relays
• Communications: RS-485 (Modbus) + Ethernet options

Industrial Features:
• Complies with API 670 standards
• -20°C to +70°C operation
• Panel-mounted IP20 enclosure

Typical Deployment:
Ideal for turbine/compressor monitoring in:

Power generation

Oil & gas

Heavy industrial applications

Configuration:
Programmable via front panel or CMS 2000 software

open
TEL:+86 13600179521
Mail:info@hongkongeasy.com
info@hongkongeasy.com
WhatsApp
Parameters
Request a Quote
Return to List
General Description
The EPRO MMS6410 is a 4-channel monitoring module designed for vibration, displacement, speed, and temperature measurement in critical industrial machinery (turbines, compressors, pumps, etc.). It provides real-time monitoring, alarm triggering, and analog/digital outputs for seamless integration with control systems.

Key Parameters
Input Specifications
Channels: 4 independent monitoring channels (configurable per parameter)

Compatible Sensors:

Eddy current probes (e.g., EPRO PR6423, PR6426)

Velocity/acceleration sensors (e.g., seismic transducers)

Speed/Tachometer signals (pulse input)

Temperature (RTD/thermocouples) (optional)

Input Signal Range:

Displacement (Proximity Probes): -2 to -20V DC

Vibration (Velocity): 0.5–20V peak

Speed: 5–15V pulse (magnetic pickup)

Measurement Capabilities
Vibration Monitoring:

Range: 0–500 µm pk-pk (adjustable)

Frequency Band: 10 Hz–1 kHz (ISO 10816 compliant)

Displacement Monitoring:

Range: 0–2.0 mm (configurable)

Accuracy: ±1% of full scale

Speed Monitoring:

Range: 0–20,000 RPM

Resolution: 1 RPM

Temperature Monitoring (optional):

RTD (Pt100) or Thermocouples (J/K type)

Outputs & Alarms
Analog Outputs: 4 channels, 4–20 mA (isolated, configurable per parameter)

Relay Outputs: 8 programmable relays (SPDT, 250V AC/2A) for alarms (Danger/Warning)

Communication:

RS-485 (Modbus RTU)

Optional Ethernet or Profibus (depending on variant)

Power & Environmental
Power Supply: 24V DC (±10%) or 100–240V AC

Operating Temperature: -20°C to +70°C

Enclosure Rating: IP20 (panel-mounted)

Display & Interface
Front Panel: LCD with real-time values, bargraphs, and alarm status

Configuration: Keypad or software (EPRO CMS 2000 configuration tool)

Typical Applications
Turbine/compressor protection systems

Bearing health & shaft displacement monitoring

Integration with DCS/PLC systems

Compliance & Certifications
Standards: API 670, ISO 10816

Certifications: CE, ATEX (optional)